Sign in

No products in the cart.

Mo-Fr 8am 6pm CT

Additel 875 - Dry Well Calibrator

Price range: $7,710.00 through $17,149.00

Do you need more information? Contact us
Overview
Description
Documents
Accessories
Applications

Features

  • Three models ranging from -40°C to 660°C: 875-155, 875-350, 875-660
  • Dual-zone control for axial uniformity
  • Stability (30 min) from ±0.01°C at full range (875-155)
  • Color touch screen display, 6.5 in (165 mm)
  • Process Calibrator option (ADT875PC): reference thermometer input plus two channels for RTD, TC, mA, V, and switch
  • Full HART field communicator (PC option)
  • Set point control by internal sensor or external reference PRT
  • Self-calibration of the internal control sensor against your external reference
  • Choose Your Own Range (CYOR) option: calibration optimized over your selected range
  • USB A, USB B, RJ45, WiFi, Bluetooth communication

Applications

  • Field calibration of RTDs, thermocouples, and temperature switches
  • Calibration of temperature transmitters, including HART devices, with the PC option
  • Comparison calibration against a reference PRT in the lab
  • Documented calibration tasks with as found and as left data

Technical Specifications

  • Temperature range at 23°C: -40°C to 155°C (875-155); 33°C to 350°C (875-350); 33°C to 660°C (875-660)
  • Display accuracy: ±0.18°C at full range (875-155); ±0.2°C at full range (875-350); ±0.3°C at 33°C, ±0.3°C at 420°C, ±0.5°C at 660°C (875-660)
  • Stability (30 min): ±0.01°C at full range (875-155); ±0.02°C at full range (875-350); ±0.02°C at 33°C, ±0.03°C at 50°C, ±0.04°C at 420°C, ±0.04°C at 660°C (875-660)
  • Axial uniformity at 60 mm (2.4 in): ±0.07°C at full range (875-155); ±0.04°C at 33°C, ±0.1°C at 200°C, ±0.2°C at 350°C (875-350); ±0.05°C at 33°C, ±0.3°C at 420°C, ±0.5°C at 660°C (875-660)
  • Radial uniformity: ±0.01°C at full range (875-155); ±0.01°C at 33°C, ±0.015°C at 200°C, ±0.02°C at 350°C (875-350); ±0.02°C at 33°C, ±0.05°C at 420°C, ±0.1°C at 660°C (875-660)
  • Loading effect: ±0.1°C display sensor, ±0.02°C external sensor (875-155); ±0.15°C display sensor, ±0.015°C external sensor (875-350); ±0.15°C display sensor, ±0.035°C external sensor (875-660)
  • Hysteresis (display sensor): 0.025°C (875-155); 0.03°C (875-350); 0.1°C (875-660)
  • Immersion depth: 145 mm (5.70 in) (875-155); 150 mm (5.90 in) (875-350, 875-660)
  • Insert OD: 25.8 mm (1.02 in) (875-155); 24.8 mm (0.98 in) (875-350, 875-660)
  • Heating time: 5 min -40°C to 23°C, 8 min 23°C to 155°C (875-155); 10 min 33°C to 350°C (875-350); 15 min 33°C to 660°C (875-660)
  • Typical time to stability: 10 min
  • Resolution: 0.01°C; units: °C, °F, and K
  • Environmental conditions: 8°C to 38°C guaranteed accuracy; 0°C to 50°C, 0% to 90% RH non-condensing, 3000 m altitude for normal operation
  • Size (H x W x D): 320 x 170 x 330 mm (12.6 x 6.7 x 13.0 in)
  • Weight: 9.9 kg (21.8 lbs) (875-155); 8.6 kg (18.9 lbs) (875-350, 875-660)
  • Power requirements: 90-254 VAC, 45-65 Hz, 580 W (875-155); 90-254 VAC, 45-65 Hz, 1200 W (875-350, 875-660)
  • PC option, reference readout accuracy for 100 ohm PRT (probe not included): ±0.009°C at -40°C, ±0.010°C at 0°C, ±0.017°C at 155°C, ±0.026°C at 350°C, ±0.042°C at 660°C
  • PC option, RTD channels: 2, 2-, 3-, or 4-wire; 25 Ω to 400 Ω: ±0.008% RD
  • PC option, TC channels: 2, types S, R, K, B, N, E, J, T, C, D, G, L, and U; TC voltage accuracy 0.02% RD + 5 µV
  • PC option, current: -30 mA to 30 mA, 0.02% RD + 2 µA; voltage: -12 V to 12 V and -30 V to 30 V, ±0.02% RD + 2 mV; DC 24V output 24V ±10%, max 60 mA
  • PC option, documentation: up to 1,000 tasks, each storing up to 10 results with as found and as left data
  • Communication: USB A, USB B, RJ45, WiFi, Bluetooth
  • Warranty: 1 year

A good dry well is more than a stable heat source. Field calibration also needs radial and axial uniformity you can trust, low loading effect with several probes in the insert, and fast heating and cooling between points. The Additel 875 delivers that in a rugged, portable unit with dual-zone control and a 6.5 in color touch screen, across three models from -40°C to 660°C.

The Process Calibrator option turns the 875 into a complete temperature calibration station: a reference PRT input, two device channels that read RTDs, thermocouples, mA, V, and switches, 24V loop power, and a full HART communicator. It stores up to 1,000 documented tasks with as found and as left results, and can control the set point by the external reference PRT. A self-calibration function checks the internal control sensor against your reference, and the Choose Your Own Range option optimizes calibration over the range you actually use.

Pair the 875 with AccuMac AM17XX secondary reference PRTs, read probes in the lab with the Additel 282, and use the 9080 thermocouple cable kit for TC work. For higher temperatures and reference-level uniformity, see the Additel 878.

AV-TEK Consulting supplies Additel temperature calibration equipment across North America and helps you choose the model, insert, and options for your work. Browse all temperature calibrators and furnaces or the full Additel range.

Manuals

Links

Spec Sheets

Additel_ADT875_Datasheet
Professional Brands
Quality and Safety
Fast shipping
Best couriers
Check your order
In your Account
Secure Payments