Additel AccuMac AM17XX - Secondary Reference PRTs

Price range: $1,275.00 through $3,417.00

Do you need more information? Contact us
Overview
Description
Documents
Accessories
Applications

Features

  • Three models: AM1710 (-40°C to 160°C), AM1730 (-200°C to 420°C), AM1751 (-200°C to 670°C)
  • Calibrated accuracy (k=2): ±0.015°C at 0.01°C on all three models
  • Short-term stability: <0.007°C; long term drift at 0.01°C: <0.01°C at TPW after 100 hours at maximum temperature
  • Hysteresis: ≤0.005°C; thermal shock: <0.005°C after 10 thermal cycles
  • Fully meets the ITS-90 criteria for reference thermometers
  • Straight and 90° bend versions; 1/4 in and 3/16 in sheath diameters depending on model
  • Additel smart connector (Lemo) with 4-lead, 2 m cable
  • ISO 17025 accredited calibration included

Applications

  • Reference probe for the Additel 878 and 875 dry wells and the 835 portable calibration bath
  • Reference thermometry with the Additel 282 dual-channel readout
  • Comparison calibration of RTDs, PRTs and thermocouples in the laboratory
  • On-site verification of process temperature sensors against a traceable reference

Technical Specifications

  • Nominal resistance at 0.01°C: 100 Ω
  • Temperature coefficient: 0.003925 Ω/Ω/°C
  • Short term stability: <0.007°C
  • Thermal shock: <0.005°C after 10 thermal cycles from minimum to maximum temperatures
  • Hysteresis: ≤0.005°C
  • Self-heating: 50 mW/°C
  • Measurement current: 0.5 mA or 1 mA (AM1730, AM1751)
  • Sensor element length: 1.26 in (32 mm); sensor location: 0.12 in (3 mm) from tip
  • Insulation resistance: >1000 MΩ at room temperature
  • External leads: Teflon™ insulated copper wire, 4 leads, 6.5 feet (2 meters)
  • Connection: Additel smart connector (ADT versions)
  • Calibration: ISO 17025 accredited calibration
  • Calibrated accuracy includes calibration and 100 hour drift; probe calibration ranges may differ from probe temperature ranges

Model List

  • AM1710: -40°C to 160°C; calibrated accuracy (k=2) ±0.025°C at -40°C, ±0.015°C at 0.01°C, ±0.025°C at 160°C; long term drift <0.01°C at TPW after 100 hours at 160°C; response time 9 seconds; stainless steel sheath, 0.25 in (6.35 mm) diameter; handle 0.59 in (15 mm) OD x 2.56 in (65 mm); handle temperature range -40°C to 160°C
  • AM1710-12: sheath length 12 in (305 mm)
  • AM1710-BEND: 7 in (178 mm) from tip to bend, 4 in (101.6 mm) from bend to handle
  • AM1710-6 and AM1710-9: 3/16 in x 6 in and 3/16 in x 9 in (Additel price list)
  • AM1730: -200°C to 420°C; calibrated accuracy (k=2) ±0.025°C at -40°C, ±0.015°C at 0.01°C, ±0.035°C at 420°C; long term drift <0.01°C at TPW after 100 hours at max temperature; Inconel™ sheath; handle temperature range -50°C to 180°C
  • AM1730-12: 0.25 in (6.35 mm) OD x 12 in (305 mm); response time 9 seconds; handle 0.59 in (15 mm) OD x 2.56 in (65 mm)
  • AM1730-9: 0.188 in (4.775 mm) OD x 9 in (228.6 mm); response time 6 seconds; handle 0.39 in (10 mm) OD x 1.97 in (50 mm)
  • AM1730-BEND: 0.25 in (6.35 mm) OD x 9.75 in (247.65 mm) vertical x 4.5 in (114.3 mm) horizontal; response time 9 seconds
  • AM1751: -200°C to 670°C; calibrated accuracy (k=2) ±0.025°C at -40°C, ±0.015°C at 0.01°C, ±0.035°C at 420°C, ±0.05°C at 661°C; long term drift <0.01°C at TPW after 100 hours at 661°C; response time 9 seconds; Inconel™ sheath; handle 0.59 in (15 mm) OD x 2.56 in (65 mm); handle temperature range -50°C to 180°C
  • AM1751-12: 0.25 in (6.35 mm) OD x 12 in (305 mm)
  • AM1751-BEND: 0.25 in (6.35 mm) OD x 9.75 in (247.65 mm) vertical x 4.5 in (114.3 mm) horizontal

A reference probe sets the floor of every temperature calibration: each sensor checked against it inherits its error. The AccuMac AM17XX series is built to keep that error small and documented, with a sensing element designed to protect the platinum wire from contamination at high temperature, a support structure balanced for hysteresis and shock, and an ISO 17025 accredited calibration on every probe.

The model follows the range. The AM1710 covers -40°C to 160°C with a stainless steel sheath; the AM1730 extends to 420°C and the AM1751 to 670°C, both with an Inconel™ sheath. Straight probes suit wells and baths, and a 90° bend version is available on each model.

With the Additel smart connector the probes work with the Additel 878 reference dry well, the Additel 875 dry block calibrator, the Additel 835 portable calibration bath and the Additel 282 dual-channel reference thermometer. For industrial grade references see the AccuMac AM16XX precision industrial PRTs; for tighter uncertainty, the AccuMac AM176X secondary SPRTs.

AV-TEK Consulting supplies Additel temperature calibration equipment across North America and helps you match reference probe, readout and heat source to your uncertainty budget. Browse all thermometer readouts and probes or the full Additel range.

Manuals

Links

Spec Sheets

Additel_AM1710_Datasheet
Additel_AM1730_Datasheet
Additel_AM1751_Datasheet
Professional Brands
Quality and Safety
Fast shipping
Best couriers
Check your order
In your Account
Secure Payments
Choose what you prefer

AV-TEK / Houston Office
4123 Hollister, Suite G
Houston, TX, 77080

AV-TEK / San Marcos Office
900 Barnes Dr. Suite 200
San Marcos, TX, 78666

Subscribe to AV-TEK

Receive the latest news, updates and special offers right to your inbox.
© 2026 AV-TEK
phone-handsetchevron-downcross-circle