Description
Documents
Accessories
Applications

Features

  • Extended temperature range: -30°C to 250°C across two models
  • Display accuracy ± 0.1 °C; EXT REF accuracy ± 0.05 °C with AM1730-12-ADT
  • Dual-zone heating: an external heater near the vapor chamber plus an internal heater in the fluid, for stability in changing field conditions
  • Opening size 90 x 90 mm, maximum media depth 155 mm, for probes of many types and sizes
  • Optional built-in three-channel process calibrator (PC): thermometer, multimeter, 24V power supply, and HART communicator
  • One touch automated self-calibration of the internal reference sensor
  • Remote control by Bluetooth or WiFi hotspot (point-to-point) or WiFi and LAN (cloud mode)
  • Integrated overflow tank, drain port, and sealed top cover with a hand-tight screw for transport
  • 6.5 in color touch screen

Applications

  • On-site calibration of sanitary, short-stem, and special shape temperature sensors in pharmaceutical, food, and biotech plants
  • Batch calibration of RTDs, thermocouples, and temperature switches
  • Calibration of temperature transmitters and HART temperature transmitters with the PC option
  • Field and mobile calibration where a liquid bath gives better uniformity than a dry well
  • Quality control and maintenance shops that need a portable reference bath

Technical Specifications

  • Models: 835-150, 835-250 (non-PC and PC versions)
  • Range at 23°C: -30 ~ 150 °C (835-150); 33 ~ 250 °C (835-250)
  • Display accuracy: ± 0.1 °C
  • EXT REF accuracy: ± 0.05 °C (with AM1730-12-ADT)
  • Stability (10 mins): ± 0.01 °C (835-150); ± 0.015 °C (835-250)
  • Uniformity: ± 0.02 °C (835-150); ± 0.03 °C (≤200 °C), ± 0.04 °C (> 200 °C) (835-250)
  • Repeatability: 0.04 °C (internal reference); 0.01 °C (external reference)
  • Heating time, 835-150: 18 min (-30 ~ 23 °C), 30 min (23 ~ 150 °C)
  • Heating time, 835-250: 20 min (23 ~ 100 °C), 35 min (100 ~ 250 °C)
  • Cooling time, 835-150: 45 min (150 ~ 23 °C), 45 min (23 ~ -25 °C), 55 min (23 ~ -30 °C)
  • Cooling time, 835-250: 30 min (250°C ~ 100°C), 45 min (100°C ~ 33°C)
  • Typical time to stability: 10 min
  • Working area: φ75mm circular zone of bath center, 15mm above bath bottom and 65mm below the fluid level
  • Opening size: 90 x 90 mm; maximum media depth: 155 mm; volume: Max 2.5 L
  • Media recommendation: 10cSt DOW CORNING XIAMETER PMX-200 (835-150); 50cSt DOW CORNING XIAMETER PMX-200 (835-250)
  • Resolution: adjustable, up to 0.001 °C; units °C, °F, K
  • Display: 6.5 in (165 mm) color touch screen, 640*480 pixels
  • Size: 7.5 × 16.2 × 15.2 in (191 W × 412 H × 387 D mm); 7.5 × 18.1 × 16.77 in (191 W × 460 H × 426 D mm) with handle, overflow tube
  • Weight, 835-150: 35.27 lbs (16 kg) non-PC; 35.94 lbs (16.3 kg) PC
  • Weight, 835-250: 30.86 lbs (14 kg) non-PC; 31.53 lbs (14.3 kg) PC
  • Power requirements: -110V: 100~120V, 50/60 Hz, fuse 8A 250V; -220V: 200~230V, 50/60 Hz, fuse 4A 250V; maximum power consumption 800W
  • Operating temperature: 0 ~ 40°C; guaranteed accuracy: 13 ~ 33°C
  • IP rating: IP20, indoor use only
  • Communication: USB-A, USB-B, LAN, WIFI, BLE
  • Warranty: 1 year

Process Calibrator (PC) Option

  • Reference channel (EXT.REF): Pt25, Pt100, 4-wire, 0~400 Ω; accuracy ± 1.25 mΩ @ (0~50)Ω, ± 0.0025%RD @ (50~400)Ω
  • DUT channels RTD: Pt10, Pt25, Pt50, Pt100, Pt200, Pt500, Pt1000, Cu10, Cu50, Cu100, Ni100, Ni120; 2/3/4-wire; 0~400 Ω and 0~4000 Ω, auto switch
  • DUT channels TC: S, R, B, K, N, E, J, T, C, D, G, L, U; mV range -75~75 mV; accuracy ± (0.01%RD + 5 μV); cold junction accuracy ± 0.2 °C
  • Current measurement: -30~30 mA, accuracy ± (0.01% RD + 2 μA)
  • Voltage measurement: -12~12 V and -30~30 V, accuracy ± (0.01% RD+0.6 mV)
  • Loop power: 24 V, max 60mA
  • Connections: Lemo connector, quick push connectors, and φ4mm banana jacks

Many temperature sensors in pharmaceutical, food, and biotech plants are short, sanitary, or oddly shaped. They do not fit a dry well insert, and removing them for a lab calibration means downtime. A liquid bath solves the fit problem, but most baths are too heavy and slow to take into the field. The Additel 835 is built to close that gap.

The 90 x 90 mm opening and 155 mm depth accept probes of different sizes at the same time. Dual-zone heating keeps the fluid stable when the environment changes, with stability down to ± 0.01 °C on the 835-150. An automated self-calibration routine checks the internal reference sensor without manual calculations. The 835-150 covers -30 to 150 °C, the 835-250 covers 33 to 250 °C.

With the PC option, the bath becomes a complete field station: a reference channel for an AccuMac AM17XX secondary reference PRT and two DUT channels for RTDs, thermocouples, transmitters, and HART devices. For dry block work, see the Additel 875; for a standalone readout, the Additel 282.

AV-TEK Consulting supplies Additel temperature calibration equipment across North America and helps you choose the model, voltage, and reference probe for your sensors. Browse all temperature calibrators and furnaces or the full Additel range.

Manuals

Additel 835_Quick Start Guide

Links

Additel 835 - Overview

Spec Sheets

Additel_ADT835_Datasheet
Professional Brands
Quality and Safety
Fast shipping
Best couriers
Check your order
In your Account
Secure Payments